Short Communication

Split Viewer

Korean J. Vet. Serv. 2022; 45(3): 243-248

Published online September 30, 2022

https://doi.org/10.7853/kjvs.2022.45.3.243

© The Korean Socitety of Veterinary Service

A HPLC-UV method for quantification of ivermectin in solution from veterinary drug products

Young-Wook Kim 1*, Wooseog Jeong 2

1Bacterial Disease Division, Animal and Plant Quarantine Agency, Gimcheon 39660, Korea
2Veterinary Drugs & Biologics Division, Animal and Plant Quarantine Agency, Gimcheon 39660, Korea

Correspondence to : Young-Wook Kim
E-mail: kimyu1223@korea.kr
https://orcid.org/0000-0003-0498-6594

Received: August 2, 2022; Revised: August 30, 2022; Accepted: September 2, 2022

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0). which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

The HPLC conditions for analysis of ivermectin in solutions dosage forms of commercial anthelmintics are different for each product. The purpose of this study was to establish a standardized chromatographic method for the quantification of ivermectin in solution. The separation was achieved on Waters Xbridge C18 column (4.6×150 nm, 5 μm) using different kinds of mobile phase composed of water/methanol/acetonitrile (15/34/51, v/v and 19.5/27.5/53, v/v), with UV detection at wavelengths 245 nm and 254 nm. A total of five commercial ivermectin in solution samples were analyzed. In this study, the optimal chromatographic conditions for analysis of ivermectin in solution were mobile phase of water/methanol/acetonitrile (15/34/51, v/v) at a flow rate of 1.0 mL/min and a detection wavelength of 245 nm using a Waters Xbridge C18 column (4.6×250 nm, 5 μm) at a column temperature of 25℃. The linearity was observed in the concentration range of 50∼150 μg/mL, with a correlation coefficient, r2=0.99999. The limit of detection and the limit of quantification were 0.88 and 2.68 μg/mL, respectively. The accuracy (% recovery) was found to be 98.9 to 100.3%. Intraday and Intermediate precisions with relative standard deviations were less than 1.0%. The content of ivermectin for five market samples ranged 91.2∼102.7%. The proposed method was also found to be robust, therefore, the method can be used for the routine analysis of ivermectin in solutions dosage forms.

Keywords Ivermectin, Veterinary, HPLC, Validation

Ivermectin (5-O-dimethyl-22,23-dihydroaverrmectin) is an antiparasitic and anthelmintic agent that kills various parasites such as lice, mange and wireworms etc (Waldia et al, 2008; Patel et al, 2015). It has been used in domestic animals for treatment of endoparasites and ectoparasites (Bark et al, 2007; Rao et al, 2017). Many analytical methods including thin-layer chromatography (TLC), high performance liquid chromatography (HPLC), high performance liquid chromatography with mass-spectrometry (HPLC-MS) or tandem mass spectrometry (MS/MS) have been developed for the analysis of ivermectin (Waldia et al, 2008). Ivermectin can be administered orally, subcutaneously, intramusculary and applied to the skin (Bark et al, 2007). Ivermectin in solution dosage form applied to the skin is easy to use in livestock farms and its use is increasing day by day (Bark et al, 2007). Currently, there are 11 ivermectin topical solutions from 11 companies registered with the Animal Drug Associations (based on sales performance from 2010 to 2019). All 11 products use HPLC to quantify ivermectin. However, the HPLC conditions such as mobile phase, flow rate and wavelength are different for each product, quality control is not performed with the same test method.

Therefore, we aimed to establish a standardized chromatographic method for the quantification of ivermectin in solution and validate the established assay.

HPLC apparatus

Agilent 1260 Infinity II series HPLC system equipped with a model G7111A pump, a model G7129A autosampler, a model G7117C UV-detector and Chemstation Software (Version Rev C.01.10 (236)) were used.

Reagents and chemicals

Ivermectin was purchased from the United States Pharmacopeia (USP) (Rockville, MD, USA), with 90.8% of purity. HPLC grade methanol was purchased from J. T. Baker Inc. (USA), Acetonitrile from Merck Ltd. (Germany). HPLC grade water was used from Fisher Scientific co. (Korea).

Preparation of standard solutions

Stock solution for ivermectin was prepared in mobile phase to get concentration of 1,000 μg/mL. Ivermectin standard working solution was prepared in mobile phase at a final concentration of 100 μg/mL. The stock solution was diluted with mobile phase into a series of standard solutions (50, 80, 100, 120 and 150 μg/mL).

Preparation of samples

Among eleven ivermectin in solutions dosage forms registered with the Animal Drug Association, five commercially available anthelmintic formulations (5 solutions) were selected and purchased except for six formulations (solutions) which production was stopped. Five commercial anthelmintic formulations contain 5 g ivermectin per liter of product. 1 mL of solutions were transferred into a 50 mL volumetric flask each, and filled with mobile phase to obtain a concentration of 100 μg/mL. These samples were sonicated for 15 min and then the samples were filtered using 0.45 μm membrane.

Chromatographic conditions and HPLC method optimization

Chromatographic separation was achieved on two mobile phases - water/methanol/acetonitrile (15:34:51, v/v) and water/methanol/acetonitrile (19.5:27.5:53, v/v) using Waters X Bridge C18 column (4.6×150 nm, 5 μm) at a flow rate of 1.0 mL/min, with UV detection at wavelengths 245 nm and 254 nm and a column temperature of 25℃.

System suitability parameters

System suitability test parameters such as peak area (mAU), retention time (Rt), number of theoretical plates (N) and tailing factor (T) for working solution were determined.

Analytical method validation

We determined parameters for method validation, including system suitability, linearity, limit of Detection (LOD), limit of quantification (LOQ), precision, accuracy and robustness according to the International Conference on Harmonization (ICH) Guidelines. System suitability was determined by six replicates of standard working solution and evaluated by calculating of the relative standard deviation (RSD) of the peak area, retention time, theoretical plates and tailing factor. Linearity was calculated by plotting peak areas versus five different concentrations (50, 80, 100, 120 and 150 ug/mL) and found out a linear equation with correlation coefficient (R2). LOD and LOQ were calculated to determine the sensitivity of the method. Precision was evaluated for repeatability and intermediate precision. Repeatability was determined by analyzing six replicates of ivermectin standard solution in the range 50, 100 and 150 μg/mL for ivermectin. Intermediate precision was determined by analyzing six replicates of ivermectin standard solution in the range 80, 120 and 120 μg/mL for ivermectin on different days. The precision of an analytical procedure was expressed as RSD %. Accuracy was determined by calculating recovery of ivermectin and conducted by adding known amounts of ivermectin standard to the in three replicates. The percent recovery was calculated by the following formula: recovery (%)= (experimental concentration/actual concentration)×100. The robustness was determined by assaying six replicates of standard solution (100 μg/mL of ivermectin) in minor changes of method parameters such as flow rate, column temperature, wavelength and mobile phase ratio.

HPLC chromatographic conditions optimization

Under the attempted chromatographic conditions, the correlation coefficients values were 0.9997 or greater. The concentrations of ivermectin were close to those when tested according to the standard test methods for products, and were within the acceptable ranges (from 90 to 120%) of the labeled amount. The results show that both wavelengths (245 nm and 254 nm) can be used to measure the amount of ivermectin, but at a wavelength of 254 nm, the peak area value is smaller for the same concentration; the peak shape obtained at 245 nm was better. In the case of the mobile phase, the C18 column (4.6×150 nm, 5 μm) was used and the retention time was shorter (16.3 to 16.5 min) for the composition of water/methanol/acetonitrile (15/34/51, v/v), than that of the composition of water/methanol/acetonitrile (19.5/27.5/53, v/v) (26.9 to 27.1 min); consequently, the analysis could be conducted faster (Fig. 1). Therefore, the optimized HPLC chromatographic conditions for the analysis of ivermectin in solution were as follows: mobile phase of water/methanol/acetonitrile (15:34:51, v/v) at a flow rate of 1.0 mL/min and a detection wavelength of 245 nm using a Waters XBridge C18 column (4.6×150 nm, 5 μm) at a column temperature of 25℃ (Table 1).

Table 1 . Optimized chromatographic conditions

ParametersChromatographic conditions
ColumnWaters X bridge C 18 4.6×150 mm, 5 μm
Mobile phaseWater:Methanol:Acetonitrile (15:34:51)
Wavelength245 nm
Column oven25℃
Flow rate1.0 mL/min
Injection volume20 μL
apparatusAgilent 1260 Infinity II series, USA


Fig. 1.Chromatograms for 100 ug/mL ivermectin standard solution: (A) Water:MEOH:ACN (15:34:51), 245 nm; (B) Water:MEOH:ACN (19.5:27.5:53), 245 nm; (C) Water: MEOH:ACN (15:34:51), 254 nm; (D) Water:MEOH:ACN (19.5:27.5: 53), 254 nm.

The optimization of chromatographic conditions led to more efficient and faster separation of ivermectin from commercial ivermectin topical solutions with different HPLC conditions.

System suitability

RSD % of peak areas, retention times and tailing factors were less than 1.0. The theoretical plates were >10,000 (Table 2). The values of system suitability parameters were all within the acceptable limits and indicated that the system was suitable for the quantification of ivermectin.

Table 2 . System suitability tests and chromatographic conditions

ParametersValue*RSD (%)Acceptable limit
Retention time (min)16.4050.09RSD ≤2
Peak area34830.1RSD ≤2
No. of theoretical plates106910.48>2000
Tailing factor1.0200.52≤2

*Mean of six values.



Linearity

The calibration curve was obtained in the concentration range of 50∼150 μg/mL. The linear regression equation for ivermectin was Y=34.98032x−4.69585 and the correlation coefficient (R2) was 0.99999 (Fig. 2). The correlation coefficient was satisfactory (R2 >0.999) and indicated that this method was reliable for quantitative detection.

Fig. 2.Calibration graph for ivermectin standard.

LOD and LOQ

Limit of detection (LOD) is the lowest concentration of analyte that can be detected, and limit of quantification (LOQ) is the lowest concentration of analyte that can be quantified with acceptable accuracy and precision. The LOD and LOQ for ivermectin were 0.88 μg/mL and 2.68 μg/mL, respectively (Table 3). The results showed the good sensitivity of the proposed method.

Table 3 . Summary of validation parameters

ParametersIvermectin
Linear range (μg/mL)50∼150
Correlation coefficient (r2)0.99999
LOD (μg/mL)0.88
LOQ (μg/mL)2.68
Intra-day precision (repeatability, RSD %)0.09∼0.17
Intermediate precision (inter-day precision, RSD %)0.07∼0.20
Accuracy (% Recovery)98.9∼100.3


Precision and accuracy

For the accuracy of the method, a recovery test was conducted by adding known amounts of ivermectin standard (80, 100 and 120 μg/mL) to the sample (19.66 μg/mL of ivermectin) and analyzing three replicates. To determine the precision of the method, repeatability (intra-day precision) and intermediate precision (inter-day precision) were evaluated. In this study, the mean recovery rate of ivermectin was 98.9% to 100.3% and the % RSD was not >2% (Table 3), indicating that the method was capable of showing good accuracy and precision.

Robustness

The values of tailing factor, theoretical plates and % RSD were within the acceptance ranges (Table 4). So, the method was shown to be robust for minor changes in method parameters such as flow rate (±0.1 mL/min), column temperature (±2℃), wavelength (±2 nm) and mobile phase composition (MEOH content ±2% v).

Table 4 . Robustness tests for Ivermectin

ParametersVariationTailing factorTheoretical plates


Value*RSD (%)Value*RSD (%)
Flow rate (mL/min)0.91.130.4572250.49
1.11.130.5965770.55
Column temperature (℃)231.130.4773550.35
271.140.6367280.22
Detection wavelength (nm)2431.140.4767610.99
2471.140.4667571.00
Mobile phase composition-Methanol content (% v)321.140.3472540.89
361.130.4272540.29

*Mean of three values.



Analysis of Sample

The optimized HPLC procedure was applied for the analysis of 5 commercially available products. The contents of ivermectin were found to be in the range from 91.2% to 102.7% and were within the acceptable ranges (90∼120%) for the percent of the labeled amount (Table 5).

Table 5 . Concentration of ivermectin in marketed solutions

Sample No.Label concentration (g/L)The standard test method (μg/mL)The proposed method

Quantity in sample (μg/mL)Concentration (%)
154.994.9699.5
255.115.13102.7
354.754.7895.6
454.704.5691.2
554.805.08101.5

The optimized HPLC chromatographic method was established and validated for the assay of ivermectin in solutions. The results of this study showed that this HPLC method is simple, rapid, accurate, precise, sensitive and robust. Therefore, this method can be performed for routine quality control analysis of ivermectin from commercial ivermectin topical solutions with different HPLC conditions.

This study was supported by research funds from the Animal and Plant Quarantine Agency (APQA), Republic of Korea (Project Code No. N-1543073-2018-22-0104).

No potential conflict of interest relevant to this article was reported.

  1. Abotaleb N, Nasr T, Elsherif Z. 2017. Development and Validation of HPTLC and HPLC Methods for Simultaneous Determination of Closantel and Ivermectin in Veterinary Drug Products 9(3):135-140.
  2. Ali MM, Abdalaziz MN. 2017. Development and Validation for HPLC Method of Assay of Ivermectin and Clorsulon in Combined Pharmaceutical Dosage Form. International Journal of Homeopathy & Natural Medicines. 3(6):45-55.
    CrossRef
  3. Bark JJ, Youk JH, Kim HK, Park HW, Lee WS. 2007. Examination of ivermectin residues in raw milk after skin administration. Korean Journal of Veterinary Service. 30(3):421-428.
  4. International Conference on Harmonization. 2005. Validation Analytical Procedures:Text and Methodology Q2(R1).
  5. Kumar Devaka NVS and Madhusudhan RV. 2019. Chromatographic Quantification of Ivermectin and Pranziquantel in the Tablets Using Stability Indicating RP-HPLC Method. Pharmaceutical Scineces. 25(3):254-261.
    CrossRef
  6. Na-Bangchang K, Choemung A. 2006. High-performance liquid chromatographic method for the determination of ivermectin in plasma 37(5):848-858.
  7. Nischal K, Somshekar B, Abhilekha PM, Radhakrishna PM. 2011. A Simple RP-HPLC Method for Estimation of Triclabendazole and Ivermectin in a Pharmaceutical Suspension Dosage Form. Current Pharma Research. 1(4):306-310.
    CrossRef
  8. Oltean EG and Nica A. 2011. Development and validation of a RP-HPLC method for the quantitation studies of ivermectin in solutions dosage forms 5(2):68-70.
  9. Padivitage N, Rustum AM. 2021. Assay of Ivermectin Including Determination and Identification of Its Related Substances in Ivermectin Injectable Product by Stability0Indicating RP-HPLC Method 84:989-997.
    CrossRef
  10. Patel AK, Patel JK. 2015. Development and validation of stability indicating RP-HPLC method for estimation of ivermectin and albendazole in pharmaceutical dosage form 3(3):57-70.
  11. Phatak HM, Vaidya VV, Rajeghadge D. 2016. A Rapid High Performance Liquid Chromatography Method for Simultaneous Quantification of Praziquantel, Ivermectin and Abamectin from Veterinary Formulations: Development, Validation and Application. International Journal for Pharmaceutical Research Scholars. 5(1):57-65.
  12. Rao BS, Rao BS. HPLC determination of fenbendazole and ivermectin simultaneously in bulk and pharmaceutical dosage forms. Indo American Journal of Pharmaceutical Sciences 4(05):1194-1204.
  13. Sanjay HS, Nimita M, Dubey PK. 2012. Development and validation of analytical method for fluconazole and ivermectin in tablet formulation by using RP-HPLC. International Research Journal of Pharmacy. 3(8):257-261.
  14. Shurbaji M, et al. 2010. Development and Validation of a New HPLC-UV Method for the Simultaneous Determination of Triclabendazole and Ivermectin B1a in a Pharmaceutical Formulation. Journal of AOAC International. 93(6):1868-1873.
    Pubmed CrossRef
  15. Vegad KL, Paranjape DB, Shah DA, Patel ED, Patel KR. 2017. Development and validation of RP-HPLC method for simultaneous estimation of ivermectin and clorsulon in ivermectin injection. Indo American Journal of Pharmaceutical Research. 7(8):523-531.
  16. Waldia A, Gupta S, Nagori BP. 2008. Validated liquid chromatographic method for simultaneous estimation of albendazole and ivermectin in tablet dosage form 15:617-620.
  17. Winalasinghe RM, Zhao D, Rustum A. 2021. Development and Validation of A Stability-Indicating Reversed-Phase High-Performance Liquid Chromatography (RP-HPLC) Method for Identification, Assay, and Estimation of Related Substances of Ivermectin Drug Substance 104(6):1505-1513.
    Pubmed CrossRef

Article

Short Communication

Korean J. Vet. Serv. 2022; 45(3): 243-248

Published online September 30, 2022 https://doi.org/10.7853/kjvs.2022.45.3.243

Copyright © The Korean Socitety of Veterinary Service.

A HPLC-UV method for quantification of ivermectin in solution from veterinary drug products

Young-Wook Kim 1*, Wooseog Jeong 2

1Bacterial Disease Division, Animal and Plant Quarantine Agency, Gimcheon 39660, Korea
2Veterinary Drugs & Biologics Division, Animal and Plant Quarantine Agency, Gimcheon 39660, Korea

Correspondence to:Young-Wook Kim
E-mail: kimyu1223@korea.kr
https://orcid.org/0000-0003-0498-6594

Received: August 2, 2022; Revised: August 30, 2022; Accepted: September 2, 2022

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0). which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

The HPLC conditions for analysis of ivermectin in solutions dosage forms of commercial anthelmintics are different for each product. The purpose of this study was to establish a standardized chromatographic method for the quantification of ivermectin in solution. The separation was achieved on Waters Xbridge C18 column (4.6×150 nm, 5 μm) using different kinds of mobile phase composed of water/methanol/acetonitrile (15/34/51, v/v and 19.5/27.5/53, v/v), with UV detection at wavelengths 245 nm and 254 nm. A total of five commercial ivermectin in solution samples were analyzed. In this study, the optimal chromatographic conditions for analysis of ivermectin in solution were mobile phase of water/methanol/acetonitrile (15/34/51, v/v) at a flow rate of 1.0 mL/min and a detection wavelength of 245 nm using a Waters Xbridge C18 column (4.6×250 nm, 5 μm) at a column temperature of 25℃. The linearity was observed in the concentration range of 50∼150 μg/mL, with a correlation coefficient, r2=0.99999. The limit of detection and the limit of quantification were 0.88 and 2.68 μg/mL, respectively. The accuracy (% recovery) was found to be 98.9 to 100.3%. Intraday and Intermediate precisions with relative standard deviations were less than 1.0%. The content of ivermectin for five market samples ranged 91.2∼102.7%. The proposed method was also found to be robust, therefore, the method can be used for the routine analysis of ivermectin in solutions dosage forms.

Keywords: Ivermectin, Veterinary, HPLC, Validation

INTRODUCTION

Ivermectin (5-O-dimethyl-22,23-dihydroaverrmectin) is an antiparasitic and anthelmintic agent that kills various parasites such as lice, mange and wireworms etc (Waldia et al, 2008; Patel et al, 2015). It has been used in domestic animals for treatment of endoparasites and ectoparasites (Bark et al, 2007; Rao et al, 2017). Many analytical methods including thin-layer chromatography (TLC), high performance liquid chromatography (HPLC), high performance liquid chromatography with mass-spectrometry (HPLC-MS) or tandem mass spectrometry (MS/MS) have been developed for the analysis of ivermectin (Waldia et al, 2008). Ivermectin can be administered orally, subcutaneously, intramusculary and applied to the skin (Bark et al, 2007). Ivermectin in solution dosage form applied to the skin is easy to use in livestock farms and its use is increasing day by day (Bark et al, 2007). Currently, there are 11 ivermectin topical solutions from 11 companies registered with the Animal Drug Associations (based on sales performance from 2010 to 2019). All 11 products use HPLC to quantify ivermectin. However, the HPLC conditions such as mobile phase, flow rate and wavelength are different for each product, quality control is not performed with the same test method.

Therefore, we aimed to establish a standardized chromatographic method for the quantification of ivermectin in solution and validate the established assay.

MATERIALS AND METHODS

HPLC apparatus

Agilent 1260 Infinity II series HPLC system equipped with a model G7111A pump, a model G7129A autosampler, a model G7117C UV-detector and Chemstation Software (Version Rev C.01.10 (236)) were used.

Reagents and chemicals

Ivermectin was purchased from the United States Pharmacopeia (USP) (Rockville, MD, USA), with 90.8% of purity. HPLC grade methanol was purchased from J. T. Baker Inc. (USA), Acetonitrile from Merck Ltd. (Germany). HPLC grade water was used from Fisher Scientific co. (Korea).

Preparation of standard solutions

Stock solution for ivermectin was prepared in mobile phase to get concentration of 1,000 μg/mL. Ivermectin standard working solution was prepared in mobile phase at a final concentration of 100 μg/mL. The stock solution was diluted with mobile phase into a series of standard solutions (50, 80, 100, 120 and 150 μg/mL).

Preparation of samples

Among eleven ivermectin in solutions dosage forms registered with the Animal Drug Association, five commercially available anthelmintic formulations (5 solutions) were selected and purchased except for six formulations (solutions) which production was stopped. Five commercial anthelmintic formulations contain 5 g ivermectin per liter of product. 1 mL of solutions were transferred into a 50 mL volumetric flask each, and filled with mobile phase to obtain a concentration of 100 μg/mL. These samples were sonicated for 15 min and then the samples were filtered using 0.45 μm membrane.

Chromatographic conditions and HPLC method optimization

Chromatographic separation was achieved on two mobile phases - water/methanol/acetonitrile (15:34:51, v/v) and water/methanol/acetonitrile (19.5:27.5:53, v/v) using Waters X Bridge C18 column (4.6×150 nm, 5 μm) at a flow rate of 1.0 mL/min, with UV detection at wavelengths 245 nm and 254 nm and a column temperature of 25℃.

System suitability parameters

System suitability test parameters such as peak area (mAU), retention time (Rt), number of theoretical plates (N) and tailing factor (T) for working solution were determined.

Analytical method validation

We determined parameters for method validation, including system suitability, linearity, limit of Detection (LOD), limit of quantification (LOQ), precision, accuracy and robustness according to the International Conference on Harmonization (ICH) Guidelines. System suitability was determined by six replicates of standard working solution and evaluated by calculating of the relative standard deviation (RSD) of the peak area, retention time, theoretical plates and tailing factor. Linearity was calculated by plotting peak areas versus five different concentrations (50, 80, 100, 120 and 150 ug/mL) and found out a linear equation with correlation coefficient (R2). LOD and LOQ were calculated to determine the sensitivity of the method. Precision was evaluated for repeatability and intermediate precision. Repeatability was determined by analyzing six replicates of ivermectin standard solution in the range 50, 100 and 150 μg/mL for ivermectin. Intermediate precision was determined by analyzing six replicates of ivermectin standard solution in the range 80, 120 and 120 μg/mL for ivermectin on different days. The precision of an analytical procedure was expressed as RSD %. Accuracy was determined by calculating recovery of ivermectin and conducted by adding known amounts of ivermectin standard to the in three replicates. The percent recovery was calculated by the following formula: recovery (%)= (experimental concentration/actual concentration)×100. The robustness was determined by assaying six replicates of standard solution (100 μg/mL of ivermectin) in minor changes of method parameters such as flow rate, column temperature, wavelength and mobile phase ratio.

RESULTS AND DISCUSSION

HPLC chromatographic conditions optimization

Under the attempted chromatographic conditions, the correlation coefficients values were 0.9997 or greater. The concentrations of ivermectin were close to those when tested according to the standard test methods for products, and were within the acceptable ranges (from 90 to 120%) of the labeled amount. The results show that both wavelengths (245 nm and 254 nm) can be used to measure the amount of ivermectin, but at a wavelength of 254 nm, the peak area value is smaller for the same concentration; the peak shape obtained at 245 nm was better. In the case of the mobile phase, the C18 column (4.6×150 nm, 5 μm) was used and the retention time was shorter (16.3 to 16.5 min) for the composition of water/methanol/acetonitrile (15/34/51, v/v), than that of the composition of water/methanol/acetonitrile (19.5/27.5/53, v/v) (26.9 to 27.1 min); consequently, the analysis could be conducted faster (Fig. 1). Therefore, the optimized HPLC chromatographic conditions for the analysis of ivermectin in solution were as follows: mobile phase of water/methanol/acetonitrile (15:34:51, v/v) at a flow rate of 1.0 mL/min and a detection wavelength of 245 nm using a Waters XBridge C18 column (4.6×150 nm, 5 μm) at a column temperature of 25℃ (Table 1).

Table 1 . Optimized chromatographic conditions.

ParametersChromatographic conditions
ColumnWaters X bridge C 18 4.6×150 mm, 5 μm
Mobile phaseWater:Methanol:Acetonitrile (15:34:51)
Wavelength245 nm
Column oven25℃
Flow rate1.0 mL/min
Injection volume20 μL
apparatusAgilent 1260 Infinity II series, USA


Figure 1. Chromatograms for 100 ug/mL ivermectin standard solution: (A) Water:MEOH:ACN (15:34:51), 245 nm; (B) Water:MEOH:ACN (19.5:27.5:53), 245 nm; (C) Water: MEOH:ACN (15:34:51), 254 nm; (D) Water:MEOH:ACN (19.5:27.5: 53), 254 nm.

The optimization of chromatographic conditions led to more efficient and faster separation of ivermectin from commercial ivermectin topical solutions with different HPLC conditions.

System suitability

RSD % of peak areas, retention times and tailing factors were less than 1.0. The theoretical plates were >10,000 (Table 2). The values of system suitability parameters were all within the acceptable limits and indicated that the system was suitable for the quantification of ivermectin.

Table 2 . System suitability tests and chromatographic conditions.

ParametersValue*RSD (%)Acceptable limit
Retention time (min)16.4050.09RSD ≤2
Peak area34830.1RSD ≤2
No. of theoretical plates106910.48>2000
Tailing factor1.0200.52≤2

*Mean of six values..



Linearity

The calibration curve was obtained in the concentration range of 50∼150 μg/mL. The linear regression equation for ivermectin was Y=34.98032x−4.69585 and the correlation coefficient (R2) was 0.99999 (Fig. 2). The correlation coefficient was satisfactory (R2 >0.999) and indicated that this method was reliable for quantitative detection.

Figure 2. Calibration graph for ivermectin standard.

LOD and LOQ

Limit of detection (LOD) is the lowest concentration of analyte that can be detected, and limit of quantification (LOQ) is the lowest concentration of analyte that can be quantified with acceptable accuracy and precision. The LOD and LOQ for ivermectin were 0.88 μg/mL and 2.68 μg/mL, respectively (Table 3). The results showed the good sensitivity of the proposed method.

Table 3 . Summary of validation parameters.

ParametersIvermectin
Linear range (μg/mL)50∼150
Correlation coefficient (r2)0.99999
LOD (μg/mL)0.88
LOQ (μg/mL)2.68
Intra-day precision (repeatability, RSD %)0.09∼0.17
Intermediate precision (inter-day precision, RSD %)0.07∼0.20
Accuracy (% Recovery)98.9∼100.3


Precision and accuracy

For the accuracy of the method, a recovery test was conducted by adding known amounts of ivermectin standard (80, 100 and 120 μg/mL) to the sample (19.66 μg/mL of ivermectin) and analyzing three replicates. To determine the precision of the method, repeatability (intra-day precision) and intermediate precision (inter-day precision) were evaluated. In this study, the mean recovery rate of ivermectin was 98.9% to 100.3% and the % RSD was not >2% (Table 3), indicating that the method was capable of showing good accuracy and precision.

Robustness

The values of tailing factor, theoretical plates and % RSD were within the acceptance ranges (Table 4). So, the method was shown to be robust for minor changes in method parameters such as flow rate (±0.1 mL/min), column temperature (±2℃), wavelength (±2 nm) and mobile phase composition (MEOH content ±2% v).

Table 4 . Robustness tests for Ivermectin.

ParametersVariationTailing factorTheoretical plates


Value*RSD (%)Value*RSD (%)
Flow rate (mL/min)0.91.130.4572250.49
1.11.130.5965770.55
Column temperature (℃)231.130.4773550.35
271.140.6367280.22
Detection wavelength (nm)2431.140.4767610.99
2471.140.4667571.00
Mobile phase composition-Methanol content (% v)321.140.3472540.89
361.130.4272540.29

*Mean of three values..



Analysis of Sample

The optimized HPLC procedure was applied for the analysis of 5 commercially available products. The contents of ivermectin were found to be in the range from 91.2% to 102.7% and were within the acceptable ranges (90∼120%) for the percent of the labeled amount (Table 5).

Table 5 . Concentration of ivermectin in marketed solutions.

Sample No.Label concentration (g/L)The standard test method (μg/mL)The proposed method

Quantity in sample (μg/mL)Concentration (%)
154.994.9699.5
255.115.13102.7
354.754.7895.6
454.704.5691.2
554.805.08101.5

CONCLUSIONS

The optimized HPLC chromatographic method was established and validated for the assay of ivermectin in solutions. The results of this study showed that this HPLC method is simple, rapid, accurate, precise, sensitive and robust. Therefore, this method can be performed for routine quality control analysis of ivermectin from commercial ivermectin topical solutions with different HPLC conditions.

ACKNOWLEDGEMENTS

This study was supported by research funds from the Animal and Plant Quarantine Agency (APQA), Republic of Korea (Project Code No. N-1543073-2018-22-0104).

CONFLICT OF INTEREST

No potential conflict of interest relevant to this article was reported.

Fig 1.

Figure 1.Chromatograms for 100 ug/mL ivermectin standard solution: (A) Water:MEOH:ACN (15:34:51), 245 nm; (B) Water:MEOH:ACN (19.5:27.5:53), 245 nm; (C) Water: MEOH:ACN (15:34:51), 254 nm; (D) Water:MEOH:ACN (19.5:27.5: 53), 254 nm.
Korean Journal of Veterinary Service 2022; 45: 243-248https://doi.org/10.7853/kjvs.2022.45.3.243

Fig 2.

Figure 2.Calibration graph for ivermectin standard.
Korean Journal of Veterinary Service 2022; 45: 243-248https://doi.org/10.7853/kjvs.2022.45.3.243

Table 1 . Optimized chromatographic conditions.

ParametersChromatographic conditions
ColumnWaters X bridge C 18 4.6×150 mm, 5 μm
Mobile phaseWater:Methanol:Acetonitrile (15:34:51)
Wavelength245 nm
Column oven25℃
Flow rate1.0 mL/min
Injection volume20 μL
apparatusAgilent 1260 Infinity II series, USA

Table 2 . System suitability tests and chromatographic conditions.

ParametersValue*RSD (%)Acceptable limit
Retention time (min)16.4050.09RSD ≤2
Peak area34830.1RSD ≤2
No. of theoretical plates106910.48>2000
Tailing factor1.0200.52≤2

*Mean of six values..


Table 3 . Summary of validation parameters.

ParametersIvermectin
Linear range (μg/mL)50∼150
Correlation coefficient (r2)0.99999
LOD (μg/mL)0.88
LOQ (μg/mL)2.68
Intra-day precision (repeatability, RSD %)0.09∼0.17
Intermediate precision (inter-day precision, RSD %)0.07∼0.20
Accuracy (% Recovery)98.9∼100.3

Table 4 . Robustness tests for Ivermectin.

ParametersVariationTailing factorTheoretical plates


Value*RSD (%)Value*RSD (%)
Flow rate (mL/min)0.91.130.4572250.49
1.11.130.5965770.55
Column temperature (℃)231.130.4773550.35
271.140.6367280.22
Detection wavelength (nm)2431.140.4767610.99
2471.140.4667571.00
Mobile phase composition-Methanol content (% v)321.140.3472540.89
361.130.4272540.29

*Mean of three values..


Table 5 . Concentration of ivermectin in marketed solutions.

Sample No.Label concentration (g/L)The standard test method (μg/mL)The proposed method

Quantity in sample (μg/mL)Concentration (%)
154.994.9699.5
255.115.13102.7
354.754.7895.6
454.704.5691.2
554.805.08101.5

References

  1. Abotaleb N, Nasr T, Elsherif Z. 2017. Development and Validation of HPTLC and HPLC Methods for Simultaneous Determination of Closantel and Ivermectin in Veterinary Drug Products 9(3):135-140.
  2. Ali MM, Abdalaziz MN. 2017. Development and Validation for HPLC Method of Assay of Ivermectin and Clorsulon in Combined Pharmaceutical Dosage Form. International Journal of Homeopathy & Natural Medicines. 3(6):45-55.
    CrossRef
  3. Bark JJ, Youk JH, Kim HK, Park HW, Lee WS. 2007. Examination of ivermectin residues in raw milk after skin administration. Korean Journal of Veterinary Service. 30(3):421-428.
  4. International Conference on Harmonization. 2005. Validation Analytical Procedures:Text and Methodology Q2(R1).
  5. Kumar Devaka NVS and Madhusudhan RV. 2019. Chromatographic Quantification of Ivermectin and Pranziquantel in the Tablets Using Stability Indicating RP-HPLC Method. Pharmaceutical Scineces. 25(3):254-261.
    CrossRef
  6. Na-Bangchang K, Choemung A. 2006. High-performance liquid chromatographic method for the determination of ivermectin in plasma 37(5):848-858.
  7. Nischal K, Somshekar B, Abhilekha PM, Radhakrishna PM. 2011. A Simple RP-HPLC Method for Estimation of Triclabendazole and Ivermectin in a Pharmaceutical Suspension Dosage Form. Current Pharma Research. 1(4):306-310.
    CrossRef
  8. Oltean EG and Nica A. 2011. Development and validation of a RP-HPLC method for the quantitation studies of ivermectin in solutions dosage forms 5(2):68-70.
  9. Padivitage N, Rustum AM. 2021. Assay of Ivermectin Including Determination and Identification of Its Related Substances in Ivermectin Injectable Product by Stability0Indicating RP-HPLC Method 84:989-997.
    CrossRef
  10. Patel AK, Patel JK. 2015. Development and validation of stability indicating RP-HPLC method for estimation of ivermectin and albendazole in pharmaceutical dosage form 3(3):57-70.
  11. Phatak HM, Vaidya VV, Rajeghadge D. 2016. A Rapid High Performance Liquid Chromatography Method for Simultaneous Quantification of Praziquantel, Ivermectin and Abamectin from Veterinary Formulations: Development, Validation and Application. International Journal for Pharmaceutical Research Scholars. 5(1):57-65.
  12. Rao BS, Rao BS. HPLC determination of fenbendazole and ivermectin simultaneously in bulk and pharmaceutical dosage forms. Indo American Journal of Pharmaceutical Sciences 4(05):1194-1204.
  13. Sanjay HS, Nimita M, Dubey PK. 2012. Development and validation of analytical method for fluconazole and ivermectin in tablet formulation by using RP-HPLC. International Research Journal of Pharmacy. 3(8):257-261.
  14. Shurbaji M, et al. 2010. Development and Validation of a New HPLC-UV Method for the Simultaneous Determination of Triclabendazole and Ivermectin B1a in a Pharmaceutical Formulation. Journal of AOAC International. 93(6):1868-1873.
    Pubmed CrossRef
  15. Vegad KL, Paranjape DB, Shah DA, Patel ED, Patel KR. 2017. Development and validation of RP-HPLC method for simultaneous estimation of ivermectin and clorsulon in ivermectin injection. Indo American Journal of Pharmaceutical Research. 7(8):523-531.
  16. Waldia A, Gupta S, Nagori BP. 2008. Validated liquid chromatographic method for simultaneous estimation of albendazole and ivermectin in tablet dosage form 15:617-620.
  17. Winalasinghe RM, Zhao D, Rustum A. 2021. Development and Validation of A Stability-Indicating Reversed-Phase High-Performance Liquid Chromatography (RP-HPLC) Method for Identification, Assay, and Estimation of Related Substances of Ivermectin Drug Substance 104(6):1505-1513.
    Pubmed CrossRef